Thermal Calibration
Fixed points are most accurate devices available for defining a temperature scale. Fixed point cells are designed to realize the liquid-solid equilibrium temperatures of certain high purity metal elements, for calibration of thermometers at ITS-90 fixed points.
TCLPL has facility to calibrate DTC by primary method (fixed point cell). Apart from TCLPL offer calibration of various sensors, oven, autoclaves, chambers for temperature and humidity parameters and covering range (-) 196°C to +1200°C
We Calibrate Temperature controller/ Indicator / THERMOCOUPLES by fixed point cell for following points:
• Liquid Nitrogen (-196.075 °C) CMC: 0.0084°C
• Triple Point of Water (0.01°C) CMC: 0.0018°C
• Freezing Point of Indium (156.598°C) CMC: 0.0030°C
• Freezing Point of Tin (231.928°C) CMC: 0.0032°C
• Freezing Point of Zinc (419.527°C) CMC: 0.0046°C
• Freezing Point of Zinc (419.527°C) CMC: 0.0046°C
What We Offer
Offerings | Technical Specification | Description |
Autoclave Calibration | -20 to 1000°C | Temperature Calibration by Comparison method |
Dew point Meter Calibration | -20 to 50°C | Dew point comparison |
THERMOHYGROMETER Calibration | 5°C to 50°C and 10 to 90% RH | Temperature and Humidity by comparison |
Glass Thermometer Calibration | -50 to 250 Deg C | Temperature Calibration |
Humidity Chamber Calibration | 30 to 90%RH @ 25°C | Humidity mapping with data logger |
Incubator Calibration | -80 to 300°C | Temperature Calibration |
NON-CONTACT THERMOMETER (Thermal Gun, Pyrometer Calibration) | -20 to 1200°C | Temperature Calibration |
Oil Bath Calibration | -20 to 1200°C | Temperature, stability, and Uniformity (Axial and Radial) |
Oven Calibration | -20 to 1200°C | Temperature by comparison |
Soldering Iron | 0 to 1200 °C | Temperature Calibration |
Temperature Indicator or Controller without sensor Calibration | -200 to 850°C(RTD) -200 to 1700°C (T/C) | Temperature by comparison temperature |
Temperature Chamber/Bath Calibration | -20 to 1200°C | Temperature Mapping, Temperature Uniformity Survey, System Accuracy Test |
Temperature Indicator with Sensor Calibration | -80°C to 1200°C | Temperature by comparison temperature |
Temperature Sensor Calibration | -80°C to 1200°C | Temperature by comparison hot |
Hot Plate | Amb to 400°c | Temperature by comparison water |
Water Bath Calibration | Amb to 100°c | Temperature by comparison |